A novel tandem mass spectrometry method for first-line screening of mainly beta-thalassemia from dried blood spots

Publication date: 10 February 2017
Source:Journal of Proteomics, Volume 154
Author(s): Chaowen Yu, Shuodan Huang, Ming Wang, Juan Zhang, Hao Liu, Zhaojian Yuan, Xingbin Wang, Xiaoyan He, Jie Wang, Lin Zou
Traditional methods for thalassemia screening are time-consuming and easily affected by cell hemolysis or hemoglobin degradation in stored blood samples. Tandem mass spectrometry (MS/MS) proved to be an effective technology for sickle cell disorders (SCD) screening. Here, we developed a novel MS/MS method for β-thalassemia screening from dried blood spots (DBS). Stable isotopic-labeled peptides were used as internal standards for quantification and calculation of the α:β-globin ratios. We used the α:β-globin ratio cutoffs to differentiate between normal individuals and patients with thalassemia. About 781 patients and 300 normal individuals were analyzed. The α:β-globin ratios showed significant difference between normal and β-thalassemia patients (P<0.01), particularly when the disease was homozygous or double heterozygous with another α- or β-thalassemia mutation. In the parallel study, all cases screened for suspected thalassemia from six hundred DBS samples by using this MS/MS method were successfully confirmed by genotyping. The intra-assay and inter-assay CVs of the ratios ranged from 2.4% to 3.9% and 4.7% to 7.1%, and there was no significant sample carryover or matrix effect for this MS/MS method. Combined with SCD screening, this MS/MS method could be used as a first-line screening assay for both structural and expression abnormalities of human hemoglobin.Biological significanceTraditional methods for thalassemia screening were depending on the structural integrity of tetramers and could be affected by hemolysis and degradation of whole blood samples, especially when stored. We used proteospecific peptides produced by the tryptic digestion of each globin to evaluate the production ratio between α- and β-globin chains, which turned out to be quite stable even when stored for more than two months. Though most of the peptides were specific to α-globin or β-globin, we only chose four most informative peptides and its stable isotopic-labeled peptides as internal standards for analysis, which could obtain a high accuracy. Currently, we are the first to address the application of MS/MS for thalassemia screening, when combined with SCD screening, this MS/MS method could be used as a first-line screening assay for both structural and expression abnormalities of human hemoglobin.

Graphical abstract

image


http://ift.tt/2hu6V4H

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Αναζήτηση αυτού του ιστολογίου