A national dosimetry audit for stereotactic ablative radiotherapy in lung

S01678140.gif

Publication date: Available online 20 January 2017
Source:Radiotherapy and Oncology
Author(s): Gail Distefano, Jonny Lee, Shakardokht Jafari, Clare Gouldstone, Colin Baker, Helen Mayles, Catharine H. Clark
Background and purposeA UK national dosimetry audit was carried out to assess the accuracy of Stereotactic Ablative Body Radiotherapy (SABR) lung treatment delivery.Methods and materialsThis mail-based audit used an anthropomorphic thorax phantom containing nine alanine pellets positioned in the lung region for dosimetry, as well as EBT3 film in the axial plane for isodose comparison. Centres used their local planning protocol/technique, creating 27 SABR plans. A range of delivery techniques including conformal, volumetric modulated arc therapy (VMAT) and Cyberknife (CK) were used with six different calculation algorithms (collapsed cone, superposition, pencil-beam (PB), AAA, Acuros and Monte Carlo).ResultsThe mean difference between measured and calculated dose (excluding PB results) was 0.4±1.4% for alanine and 1.4±3.4% for film. PB differences were −6.1% and −12.9% respectively. The median of the absolute maximum isodose-to-isodose distances was 3mm (−6mm to 7mm) and 5mm (−10mm to +19mm) for the 100% and 50% isodose lines respectively.ConclusionsAlanine and film is an effective combination for verifying dosimetric and geometric accuracy. There were some differences across dose algorithms, and geometric accuracy was better for VMAT and CK compared with conformal techniques. The alanine dosimetry results showed that planned and delivered doses were within ±3.0% for 25/27 SABR plans.



http://ift.tt/2iN2aFD

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Αναζήτηση αυτού του ιστολογίου