Publication date: Available online 24 February 2017
Source:Ultrasound in Medicine & Biology
Author(s): Jung Hyun Yoon, Mi Kyung Song, Eun-Kyung Kim
We evaluated the agreement and diagnostic performance of strain ratio values using measurements made with one and two user-defined regions of interest (ROIs) on breast elastography. Two hundred forty-three breast masses of 226 women (mean age: 48.2 y) were included. Ultrasonography (US) and elastography images of the masses were recorded. Strain ratio was measured twice on the same elastography image; strain ratio 1, applying one ROI at the target mass for measurement, and strain ratio 2, applying one ROI at the target mass and another ROI as reference strain. The two strain ratio measurements were in substantial agreement, with an intra-class correlation coefficient of 0.655 (95% confidence interval: 0.577–0.722). Specificity, positive predictive value and accuracy (cutoffs: 2.66 and 2.35) were significantly improved for US combined with the two strain ratio measurements (all p values < 0.05). Strain ratios measured using one or two user-defined ROIs were in substantial agreement, both contributing to the improved diagnostic performance of breast US.