Publication date: 1 August 2017
Source:Talanta, Volume 170
Author(s): Beatriz Gómez-Nieto, Mª Jesús Gismera, Mª Teresa Sevilla, Jorgina Satrústegui, Jesús R. Procopio
A micro-sampling and straightforward method based on high resolution continuum source atomic absorption spectrometry (HR-CS AAS) was developed to determine extracellular and intracellular Ca in samples of interest in clinical and biomedical analysis. Solid sampling platforms were used to introduce the micro-samples into the graphite furnace atomizer. The secondary absorption line for Ca, located at 239.856nm, was selected to carry out the measurements. Experimental parameters such as pyrolysis and atomization temperatures and the amount of sample introduced for the measurements were optimized. Calibration was performed using aqueous standards and the approach to measure at the wings of the absorption lines was employed for the expansion of the linear response range. The limit of detection was of 0.02mgL−1 Ca (0.39ng Ca) and the upper limit of linear range was increased up to 8.0mgL−1 Ca (160ng Ca). The proposed method was used to determine Ca in mitochondrial suspensions and whole blood samples with successful results. Adequate recoveries (within 91–107%) were obtained in the tests performed for validation purposes.
Graphical abstract
http://ift.tt/2nImqHx