Ετικέτες

Πέμπτη 26 Ιανουαρίου 2017

Improving structural brain images acquired with the 3D FLASH sequence

S0730725X.gif

Publication date: May 2017
Source:Magnetic Resonance Imaging, Volume 38
Author(s): Jinghua Wang, Lili He, Hairong Zheng, Zhong-Lin Lu
The three-dimension Fast Low Angle SHot Magnetic Resonance Imaging (3D FLASH) sequence has been widely adopted in medical diagnostic imaging because of its availability, simplicity, and high spatial resolution. To improve the quality of structural brain images acquired with the 3D FLASH sequence, we developed a parameter optimization scheme and image inhomogeneity correction methods. The optimal imaging parameters were determined by maximizing gray-matter and white-matter CNR efficiency. Compared to protocols based on published parameters, applying the proposed optimal imaging parameters increased CNR efficiency by >10%. Image inhomogeneity, including signal and CNR inhomogeneity, was corrected by the choice of an optimal flip angle, estimated transmit function, and estimated receive sensitivity. As a result, our optimization and image inhomogeneity correction greatly improved the quality of images acquired with the 3D FLASH sequence.