In-line moisture monitoring in fluidized bed granulation using a novel multi-resonance microwave sensor

Publication date: 1 August 2017
Source:Talanta, Volume 170
Author(s): Johanna Peters, Kathrin Bartscher, Claas Döscher, Wolfgang Taute, Michael Höft, Reinhard Knöchel, Jörg Breitkreutz
Microwave resonance technology (MRT) is known as a process analytical technology (PAT) tool for moisture measurements in fluid-bed granulation. It offers a great potential for wet granulation processes even where the suitability of near-infrared (NIR) spectroscopy is limited, e.g. colored granules, large variations in bulk density. However, previous sensor systems operating around a single resonance frequency showed limitations above approx. 7.5% granule moisture. This paper describes the application of a novel sensor working with four resonance frequencies. In-line data of all four resonance frequencies were collected and further processed. Based on calculation of density-independent microwave moisture values multiple linear regression (MLR) models using Karl-Fischer titration (KF) as well as loss on drying (LOD) as reference methods were build. Rapid, reliable in-process moisture control (RMSEP≤0.5%) even at higher moisture contents was achieved.