Preparation of an Au-Pt alloy free from artifacts in magnetic resonance imaging

S0730725X.gif

Publication date: Available online 9 July 2017
Source:Magnetic Resonance Imaging
Author(s): Tomonobu Kodama, Ryusuke Nakai, Kenji Goto, Kunihiro Shima, Hiroo Iwata
PurposeWhen magnetic resonance imaging (MRI) is performed on patients carrying metallic implants, artifacts can disturb the images around the implants, often making it difficult to interpret them appropriately. However, metallic materials are and will be indispensable as raw materials for medical devices because of their electric conductivity, visibility under X-ray fluoroscopy, and other favorable features. What is now desired is to develop a metallic material which causes no artifacts during MRI.Materials and methodsIn the present study, we prepared a single-phase and homogeneous Au-Pt alloys (Au; diamagnetic metal, and Pt; paramagnetic metal) by the processing of thermal treatment. Volume magnetic susceptibility was measured with a SQUID Flux Meter and MRI artifact was evaluated using a 1.5-T scanner.ResultsAfter final thermal treatment, an entirely recrystallized homogeneous organization was noted. The Au-35Pt alloy was shown to have a volume magnetic susceptibility of −8.8ppm, causing almost free from artifacts during MRI.ConclusionsWe thus prepared an Au-35Pt alloy which had a magnetic susceptibility very close to that of living tissue and caused much fewer artifacts during MRI. It is promising as a material for spinal cages, intracranial electrodes, cerebral aneurysm embolization coils, markers for MRI and so on.