Publication date: March 2018
Source:Journal of Environmental Radioactivity, Volume 183
Author(s): S. Mohammad Modarresi, S. Farhad Masoudi
The relative discrepancies between the gamma spectrometry efficiency of RGU, RGTh, RGK reference materials and some soil samples have been studied using a MCNP model of a real HPGe detector. It has been shown that, in a specified geometry, efficiencies differences depend on the sample elemental composition. The elemental compositions of RGU-1, RGTh-1 reference materials and a soil sample have been evaluated using X-Ray fluorescence (XRF) method and used in the MCNP simulation along with RGK-1 and six other soil samples with different elemental compositions to calculate their efficiencies in different gamma ray energies. To estimate the maximum relative efficiencies differences between soil samples and reference materials, five soil samples with higher attenuation properties were selected from a large data set of soils elemental compositions. The results show that the efficiency differences between soil samples and reference materials are almost ignorable for more than 100 KeV gamma energies. It strongly depends on the sample attenuation factor in the lower energies, so use of a self-attenuation correction is essential for radionuclide counting in low energies gamma rays. Results show about 8 percent discrepancy between RGU and two soil samples efficiencies in 63.2 KeV energy.
http://ift.tt/2ETx86i
Medicine by Alexandros G. Sfakianakis,Anapafseos 5 Agios Nikolaos 72100 Crete Greece,00302841026182,00306932607174,alsfakia@gmail.com,
Ετικέτες
Εγγραφή σε:
Σχόλια ανάρτησης (Atom)
-
Summary Insulinomas are rare neuroendocrine tumours that classically present with fasting hypoglycaemia. This case report discusses an un...
-
The online platform for Taylor & Francis Online content New for Canadian Journal of Remote Sen...
Δεν υπάρχουν σχόλια:
Δημοσίευση σχολίου