Ετικέτες

Πέμπτη 12 Απριλίου 2018

Tricuspid stenosis: A rare and potential complication of ventricular septal occluder device

aboutbul.gif

Ganesh Kumar Munirathinam, Bhupesh Kumar, Anand Kumar Mishra

Annals of Cardiac Anaesthesia 2018 21(2):195-199

Asymmetrical septal occluder device (ASOD) has made percutaneous closure of ventricular septal defect an easy and effective management option. Although there are reports of aortic and tricuspid valvular regurgitation after deployment of ASOD, only few cases of tricuspid stenosis (TS) has been reported so far in the literature. We report a case of malaligned ASOD that occurred after successful device closure resulting in TS along with mild tricuspid and aortic regurgitation requiring surgical retrieval. Transesophageal echocardiography played crucial role in detecting the cause of tricuspid valve dysfunction besides providing continuous monitoring during the procedure. We intend to emphasize the need of echocardiographic evaluation of the tricuspid valvular apparatus and aortic valve during and after the device deployment even after the successful device closure to prevent this rare complication.

https://ift.tt/2HuKDLm